The G-MED AUREN A10 Infusion Pump is designed for controlled delivery of infusion fluids at a set rate. It is intended for use by trained healthcare professionals who require adjustable infusion settings and monitoring of the infusion process.

Contact GMED to enquire about the AUREN A10 configuration, technical specifications, availability and quotation.

Key Features

  • Adjustable Infusion Settings
    The referenced infusion-pump platform supports setting the infusion rate and infusion volume according to the selected operating mode and clinical requirements.
  • Two Infusion Modes
    The technical reference documents two operating modes: ml/h and drop/min. Confirm the modes available on the supplied AUREN A10 configuration.
  • Infusion Monitoring and Alarms
    The reference manual describes alarms for conditions such as occlusion, air bubbles, an open door, low battery and completion of infusion. Exact alarm functions should be confirmed for the AUREN A10 configuration.
  • Air-Bubble Detection
    The reference platform uses ultrasonic detection to identify air bubbles in the infusion set.
  • Occlusion Pressure Settings
    The reference manual documents high, medium and low occlusion-pressure settings.
  • Battery Backup
    The reference manual specifies an internal rechargeable battery for operation when mains power is unavailable. Actual runtime depends on the confirmed configuration and operating conditions.

Technical Specifications

Specification Reference-manual details
Product name G-MED AUREN A10 Infusion Pump
Product type Volumetric Infusion Pump
Infusion modes ml/h and drop/min
Rate range, ml/h mode 1–1,200 ml/h, in 1 ml/h increments
Rate range, drop/min mode 1–266 drops/min, in 1 drop/min increments
Rate accuracy, ml/h mode ≤ ±5% after calibration
Rate accuracy, drop/min mode ≤ ±3% after calibration
Infusion volume setting 1–9,999 ml
Total infusion volume 0–9,999.9 ml
KVO rate 1–5 ml/h
Occlusion pressure levels High, medium and low
Air-bubble detection Ultrasonic; sensitivity ≥25 μL
Power supply 100–240 V AC, 50/60 Hz
Battery Internal rechargeable lithium battery, 11.1 V, capacity ≥1,800 mAh
Battery runtime More than 4 hours at 25 ml/h under the reference manual's stated conditions
Power consumption 40 VA
Dimensions 153 × 162 × 227 mm
Net weight 1.8 kg


Intended Use

The reference manual describes use for controlled administration of drug fluids through clinically accepted IV routes under the supervision of trained healthcare professionals. The manual states that the referenced device is not intended for blood transfusion.

Confirm the approved intended use and claims for the marketed AUREN A10 product before publication.

FAQ

What is the G-MED AUREN A10 Infusion Pump?
It is GMED's named infusion-pump product. Contact GMED for details of the supplied configuration and technical specifications.

Which infusion modes are documented in the reference manual?
The reference manual documents ml/h and drop/min modes. Confirm availability on the AUREN A10 configuration.

Does the reference platform have alarms?
Yes. The manual describes alarms for conditions including occlusion, air bubbles, an open door, low battery and infusion completion. The exact AUREN A10 alarm configuration should be confirmed.

Does the reference platform include battery backup?
The manual specifies an internal rechargeable battery. Confirm the battery specification and runtime for the supplied AUREN A10 unit.

How can I request a quotation?
Select GET QUOTE to contact GMED for product information, configuration details and pricing.

Interested in the G-MED AUREN A10 Infusion Pump? Contact GMED to discuss your requirements and request a quotation.

View full details